  • (EZ-Test) Inspection of Test Force

    Check wheter the test force is approximately correct or not. 1. Prepare the test force calibration jig. 2. Insert the joint for UJ into the bo ttom... Show Detail

    • No:7705
    • Date of Issue:2022/01/20 14:21
    • Last Updated:2022/01/20 14:35
    • Category: EZ-Test Series
  • (AGS-X) Inspection of Test Force

    Check wheter the test force is approximately correct or not. (The following numbers are examples.) Use a 100N weight for 1kN to 10kN load cells and a 50N... Show Detail

    • No:8008
    • Date of Issue:2022/01/20 14:21
    • Category: AGS-X Series
  • (EZ-Test) Inspection of Stroke Speed

    1. Display the jog speed setting screen by selecting the [System] - [Jog Speed], and make a note ofthe current jog speed. 2.Set the jog speed to 6mm/min. ... Show Detail

    • No:7707
    • Date of Issue:2022/01/20 14:30
    • Category: EZ-Test Series
  • (AGS-X) Inspection of Stroke Speed

    1. Display the jog speed setting screen by selecting the [System] - [Jog Speed], and make a note ofthe current jog speed. 2.Set the jog speed to 6mm/min. ... Show Detail

    • No:8010
    • Date of Issue:2022/01/20 14:30
    • Category: AGS-X Series
  • (EZ-Test) Maintenance of EZ-Test

    1. Operating Panel [Operation Buttons]:Clean with a slighyly wet soft cloth. [VFD Display]:Clean with a slighyly wet soft cloth. 2. Main Frame [Tabl... Show Detail

    • No:7710
    • Date of Issue:2022/01/20 14:38
    • Category: EZ-Test Series
  • (AGS-X) Maintenance of AGS-X series

    1. Operating Panel [Operation Buttons]:Clean with a slighyly wet soft cloth. [VFD Display]:Clean with a slighyly wet soft cloth. 2. Main Frame [Tabl... Show Detail

    • No:8012
    • Date of Issue:2022/01/20 14:38
    • Last Updated:2022/01/26 14:45
    • Category: AGS-X Series
  • (AGS-X) Emergency (Power Failure) Measures

    If a malfunction occurs, such as the crosshead not stoppinga burning odor being emitted, or apower failure, take the following actions. Then inspect the sys... Show Detail

    • No:7930
    • Date of Issue:2022/01/25 11:25
    • Last Updated:2022/01/26 10:02
    • Category: AGS-X Series
  • (AGS-X) Attaching Load Cells

    1. Switch the AGS-Xpower OFF. 2. Remove any jigs attached to the load cell. 3. Loosen the two fastener screws on the CAL ca ble conn ector on the right sid... Show Detail

    • No:7932
    • Date of Issue:2022/01/26 09:52
    • Category: AGS-X Series
    • Brached FAQ
  • (AGS-X) Safety Systems

    Once upper and lower mechanical crosshead limits are set, this feature force stops thecrosshead by shutting OFF the motor drive power if the crosshead exceed... Show Detail

    • No:7934
    • Date of Issue:2022/01/26 14:55
    • Category: AGS-X Series
    • Brached FAQ
  • (AGS-X) Jog Operations

    [JOG UP]and [JOG DOWN] buttons move the crosshead as long as they are pressed. Thecrosshead stops when the button is released. Pressing the [HIGH SPEED] but... Show Detail

    • No:8040
    • Date of Issue:2022/01/27 14:39
    • Category: AGS-X Series