FAQ within [ HPLC ]

891-900 of Total 943

≪ 90 / 95page ≫
  • Error Code:[0x9344]

    «Main Cause» Recovery failed when resending transmission data from the instrument to the workstation «Countermeasure» Check the communication environment a... Show Detail

    • No:3406
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9341]

    «Main Cause» The buffer memory used for transmitting data from the instrument to the workstation is full. «Countermeasure» Check the communication environm... Show Detail

    • No:3405
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9321]

    «Main Cause» A liquid leak has occurred in the detector. «Countermeasure» Inspect the detector for liquid leaks, and wipe away any liquids from around the l... Show Detail

    • No:3404
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9304]

    «Main Cause» There is a fault in the non-volatile memory of the detector or a non-volatile memory write error has occurred. «Countermeasure» Perform calibra... Show Detail

    • No:3403
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x5630] WARNING LOW SET TEMP

    «Main Cause» The setting is too low, relative to room temperature. «Countermeasure» Check the oven temperature setting in instrument parameter settings. If ... Show Detail

    • No:3402
    • Date of Issue:2020/12/02 15:47
    • Category: Column Oven
  • Error Code:[0x562F] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. A broken or short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and ... Show Detail

    • No:3401
    • Date of Issue:2020/12/02 15:47
    • Category: Column Oven
  • Error Code:[0x562E] ERROR EXTERNAL INPUT

    «Main Cause» Temperature control stopped because an error signal was input via the oven external input terminal. «Countermeasure» Check the connected extern... Show Detail

    • No:3400
    • Date of Issue:2020/12/02 15:47
    • Category: Column Oven
  • Error Code:[0x562D] ERROR GAS DETECTED

    «Main Cause» Temperature control was stopped because the gas sensor inside the oven detected organic solvent leakage. «Countermeasure» (1) If an organic sol... Show Detail

    • No:3399
    • Date of Issue:2020/12/02 15:47
    • Category: Column Oven
  • Error Code:[0x562C] ERROR LEAK DETECTED

    «Main Cause» Temperature control was stopped because leakage was detected in the oven. «Countermeasure» Inspect the leakage from flow channels or other loca... Show Detail

    • No:3398
    • Date of Issue:2020/12/02 15:47
    • Category: Column Oven
  • Error Code:[0x562B] ERROR OVER T.MAX

    «Main Cause» Temperature control was stopped because the oven temperature exceeded the upper limit temperature. «Countermeasure» Set the oven temperature se... Show Detail

    • No:3397
    • Date of Issue:2020/12/02 15:47
    • Category: Column Oven