FAQ - Frequently Asked Questions
- Products Search > HPLC > LC Columns
Keyword Search
Frequently Referenced FAQ
FAQ within [ LC Columns ]
-
(LC Column) Poor reproducibility of retention time betwee...
->The composition may have changed due to volatilization of the mobile phase or sample. Reprepare and check again. ->Ensure that the column is aged su... Show Detail
- No:10683
- Date of Issue:2024/09/30 09:34
-
(LC Column) Different peak shapes and retention times
->A column with a history of analysis and a new column may not obtain the same results when analyzed under the same conditions.Check that equilibration is... Show Detail
- No:10684
- Date of Issue:2024/09/30 09:34
-
(LC Column) The back pressure of the column is higher tha...
->Make sure that the column is stored according to the instructions in the manual. ->Please check that the mobile phase containing salt such as buffer... Show Detail
- No:10685
- Date of Issue:2024/09/30 09:34
-
(LC Column) Column life is too short
Contamination of the mobile phase (particulates and bacteria) Contamination of the sample (Matrices, precipitates and particulates) fragments of equipment ... Show Detail
- No:10686
- Date of Issue:2024/09/30 09:33
- Brached FAQ
-
(LC Column) General method for washing normal phase column
->Do not flush water. Please refer to the column instruction manual for solvent types that can pass through. ->When using a non-aqueous mobile phase i... Show Detail
- No:10688
- Date of Issue:2024/09/30 09:33
-
(LC) All peaks have bad shapes.
There is a possibility that voids are formed in the column because the filler is biased due to the sudden pressure fluctuation, or the filler is dissolved du... Show Detail
- No:10689
- Date of Issue:2024/09/30 09:33
- Brached FAQ
-
(LC) High column back pressure
Matrix components in the sample may be clogged. It is recommended to pass the sample through a filter of 0.45 μm or less for HPLC analysis and 0.22 μm or les... Show Detail
- No:10690
- Date of Issue:2024/09/30 09:32
- Brached FAQ
-
(LC Column) Different peak shapes and retention times wit...
A column with a history of analysis and a new column may not yield the same results when analyzed under the same conditions. Please check that equilibration ... Show Detail
- No:10876
- Date of Issue:2024/09/30 09:32
-
(LC Column) General method for washing reversed-phase column
・Flush the solution that elutes the substance adsorbed on the column. Do not precipitate out the salt inside the column. If a salt may have been used, the co... Show Detail
- No:10878
- Date of Issue:2024/09/30 09:32
-
(LC column) Retention time shifted rapidly when the aqueo...
It is recommended to increase the concentration of organic solvents or use a column that supports 100% aqueous mobile phase. Recommended product: Shim-pack ... Show Detail
- No:10682
- Date of Issue:2024/09/30 09:32
1-10 of Total 12