FAQ - Frequently Asked Questions
- Products Search > HPLC > System
Keyword Search
Frequently Referenced FAQ
FAQ within [ System ]
-
(HPLC: Amino Acid Analysis) Shutting Down the System
• For 1 to 2 Days When a series of analyses is complete and similar analyses are scheduled in 1 to 2 days time, replace the solution inside the reaction rea... Show Detail
- No:4959
- Date of Issue:2021/02/08 18:11
- Category: Amino Acid Analysis
-
(HPLC: Amino Acid Analysis) Precautions on Handling the C...
• Parts Number: Na-type 228-18837-91: Analysis Column Shim-pack Amino-Na 228-14206-91: Ammonia Trap Column ISC-30Na • Parts Number: Li-type ... Show Detail
- No:4961
- Date of Issue:2021/02/08 18:08
- Last Updated:2021/02/17 10:02
- Category: Amino Acid Analysis
-
(HPLC: Reducing Sugar Analysis) When not using and rinsin...
Since there are sugars predisposed to forming certain metal ions and weak complexes, this method performs separation of such sugars using packing that is c... Show Detail
- No:4956
- Date of Issue:2021/02/08 14:31
- Category: Reducing Sugar Analysis
-
(HPLC: Reducing Sugar Analysis) When not using and rinsin...
• Parts Number: 228-59512-41: Shim-pack ISA-09 (250 mm x 4.0 mm I.D.) 228-59512-42: Shim-pack ISA-09(G) (50 mm x 4.0 mm I.D.) • Precautions wh... Show Detail
- No:4955
- Date of Issue:2021/02/08 14:31
- Last Updated:2021/02/17 10:01
- Category: Reducing Sugar Analysis
-
(HPLC: Organic Acid Analysis) When not using and rinsing ...
• When the analysis is complete, stop pumps as follows. 1. Set the parameters as follows Pump A: 0.2 mL/min Pump B: 0.2 mL/min CTO: OFF ... Show Detail
- No:4948
- Date of Issue:2021/02/05 15:19
- Category: Organic Acid Analysis
-
(HPLC: Organic Acid Analysis) Precautions on Handling the...
Parts Number: 228-17893-91 Analysis Column,Shim-pack SCR-102H 228-17924-91 Guard Column SCR-102H (A guard column generally withstands approximately 5... Show Detail
- No:4947
- Date of Issue:2021/02/05 15:19
- Category: Organic Acid Analysis
-
Noises or fluctuation became large (Ion Chromatograph)
Bubbles have accumulated. Pass water through the suppressor. When a large noise persists even after taking the action above, perform "Regeneration Procedure... Show Detail
- No:4386
- Date of Issue:2020/10/26 14:27
- Category: Ion Chromatograph
-
Drift does not stabilize (Ion Chromatograph)
Hydrate the suppressor and use it as usual. Delivering eluent without hydrating the suppressor may cause sodium ion to enter a section from which the ion is... Show Detail
- No:4387
- Date of Issue:2020/10/26 14:24
- Category: Ion Chromatograph
-
Regeneration Procedure of ICDS (Ion Chromatograph)
Continuing passing liquid through the suppressor without synchronizing the suppressor current with the pump causes the ion exchange membranes of the suppress... Show Detail
- No:4389
- Date of Issue:2020/10/26 14:23
- Category: Ion Chromatograph
-
Hydration Procedure of ICDS (Ion Chromatograph)
The suppressor is ready for proper operation when the ion exchange membranes and mesh absorb water and swell (hydrated). Hydrate the suppressor in the fol... Show Detail
- No:4388
- Date of Issue:2020/10/26 14:03
- Category: Ion Chromatograph
1-10 of Total 13