Replacement of Plunger Seal (HPLC)
Plunger seal parts number: Standard: 228-52711-93 Plunger seal UHP(with back up ring) Option: 228-52711-94 Plunger seal, back upring, GFP *The m... Show Detail
- No:4413
- Date of Issue:2020/12/04 16:01
- Last Updated:2021/07/12 15:58
- Category: Solvent Delivery Module
- Brached FAQ
-
Maintenance for LC-40 (Pump) series (Video)
Consumable Part No. 228-57501-92 Line Filter (LC-40D X3/XS, 40B X3) 228-35871-96 Line Filter (LC-40D/XR, 40B XR) 228-70796-42 Line Filter (LC-40D ... Show Detail
- No:4340
- Date of Issue:2020/12/03 13:55
- Category: Solvent Delivery Module
- Brached FAQ
-
Consumable Part No. 228-72570-01 Air filter 1. Remove the filter holder (A) on the right side by pressing its right and left knobs (a). 2. ... Show Detail
- No:4339
- Date of Issue:2020/12/03 13:45
- Category: Column Oven
- Brached FAQ
-
Maintenance for SIL-40 series (Video)
Consumable Parts Number:228-45707-91 SUS Filter Element « Below are the changes for SIL-40C XSi» • Ring(B) is added to S... Show Detail
- No:4320
- Date of Issue:2020/12/03 13:40
- Category: Autosampler
- Brached FAQ
-
Maintenance for SPD-M40 (Video)
Part No. 228-53924-05 Dust Filter 1. Turn OFF power to the instrument. 2. Open the filter cover (A) and then remove the filter (B). ... Show Detail
- No:4309
- Date of Issue:2020/12/03 13:04
- Last Updated:2020/12/04 14:53
- Category: Detector
- Brached FAQ
-
Maintenance for SPD-40(V) (Video)
- No:4301
- Date of Issue:2020/12/03 11:16
- Last Updated:2020/12/04 14:53
- Category: Detector
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4300
- Date of Issue:2020/12/03 11:06
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4299
- Date of Issue:2020/12/03 11:05
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ
-
Check the light level. If the light level is low even with an air cell or if the lamp has been operated for more than 2000 hours, then replace the lamp. ... Show Detail
- No:4298
- Date of Issue:2020/12/03 11:03
- Category: Detector
- Brached FAQ
-
Specify an appropriate range value. Set [LAMP] to [D2]. The light source lamp illuminates. Click the [ZERO] button in the [RUN] window. Adjust the zero p... Show Detail
- No:4297
- Date of Issue:2020/12/03 11:01
- Category: Detector
- Brached FAQ