• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ HPLC ]

881-890 of Total 943

≪ 89 / 95page ≫
  • Error Code:[0x937A]

    «Main Cause» The lamp is turned off because the temperature sensor of the detector lamp housing is malfunctioning. «Countermeasure» Press the (power switc... Show Detail

    • No:3416
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9379]

    «Main Cause» The leak sensor of the detector is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sale... Show Detail

    • No:3415
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9378]

    «Main Cause» The leak sensor of the detector is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales... Show Detail

    • No:3414
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9374]

    «Main Cause» The [Cell Temperature] settings in the method parameters are set to a value outside the setting range. «Countermeasure» Set the [Cell Temperatu... Show Detail

    • No:3413
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9363]

    «Main Cause» The temperature control cell temperature does not lower to the set temperature even after 5 minutes have elapsed. «Countermeasure» Set the [Cel... Show Detail

    • No:3412
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9361]

    «Main Cause» There is a fault in the control circuit of the temperature control cell. «Countermeasure» Replace the temperature control cell. Show Detail

    • No:3411
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9356]

    «Main Cause» The IP address of the instrument is duplicated on a different device. «Countermeasure» Change the IP addresses so there are no duplicates. If ... Show Detail

    • No:3410
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9353]

    «Main Cause» The detector cell temperature does not reach the set temperature. «Countermeasure» Replace the temperature control cell. If the error is not ... Show Detail

    • No:3409
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9352]

    «Main Cause» Communication between the instrument and the workstation was interrupted. «Countermeasure» Check the communication environment and operating st... Show Detail

    • No:3408
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9345]

    «Main Cause» Light intensity was exceeded because the detector took a long exposure time. «Countermeasure» Calibrate the exposure time. Show Detail

    • No:3407
    • Date of Issue:2020/12/02 16:09
    • Category: Detector

881-890 of Total 943

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.