Frequently Referenced FAQ

FAQ within [ HPLC ]

821-830 of Total 943

≪ 83 / 95page ≫
  • Error Code:[0x920C]

    «Main Cause» The leak sensor in the autosampler low pressure valve is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail

    • No:3476
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x920B]

    «Main Cause» The leak sensor in the autosampler low pressure valve has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn the... Show Detail

    • No:3475
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x920A]

    «Main Cause» Temperature control is turned off because the autosampler rack R heater has overheated and the thermal fuse has blown. «Countermeasure» Press t... Show Detail

    • No:3474
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9208]

    «Main Cause» Temperature control is turned off because the autosampler rack L heater has overheated and the thermal fuse has blown. «Countermeasure» Press t... Show Detail

    • No:3473
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9207]

    «Main Cause» Temperature control is turned off because voltage of the autosampler dehumidifier unit is too high. «Countermeasure» Press the (power switch) t... Show Detail

    • No:3472
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9206]

    «Main Cause» Temperature control is turned off because voltage of the autosampler rack R cooling unit is too high. «Countermeasure» Press the (power switch)... Show Detail

    • No:3471
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9204]

    «Main Cause» Temperature control is turned off because voltage of the autosampler rack L cooling unit is too high. «Countermeasure» Press the (power switch)... Show Detail

    • No:3470
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9203]

    «Main Cause» Could not detect the rotation direction home position of the autosampler measuring pump. «Countermeasure» Press the (power switch) to turn the ... Show Detail

    • No:3469
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9202]

    «Main Cause» There is a liquid leak from the high pressure valve, or from a unit positioned above the high pressure value (column oven or detector, etc.). «... Show Detail

    • No:3468
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9201]

    «Main Cause» There is a liquid leak from the low pressure valve or metering pump, or from a unit positioned above the low pressure value (high pressure valve... Show Detail

    • No:3467
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler