«Main Cause» The gas sensor of the column oven has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn the power OFF and conta... Show Detail
- No:3494
- Date of Issue:2020/12/02 16:10
- Category: Column Oven
-
«Main Cause» An error was detected with the flow channel selection valve in the column oven. «Countermeasure» Press the (power switch) to turn the power OFF... Show Detail
- No:3493
- Date of Issue:2020/12/02 16:10
- Category: Column Oven
-
«Main Cause» The column oven door is open when the column oven is above 50ºC. «Countermeasure» Clear the error. The alarm sound and error message disappear.... Show Detail
- No:3492
- Date of Issue:2020/12/02 16:10
- Category: Column Oven
-
«Main Cause» Teaching has not been executed for the autosampler injection port. «Countermeasure» Adjust the position of the needle in relation t... Show Detail
- No:3491
- Date of Issue:2020/12/02 16:10
- Last Updated:2021/01/19 12:52
- Category: Autosampler
-
«Main Cause» Could not detect the home position of the autosampler high pressure valve. «Countermeasure» Press the (power switch) to turn the power OFF and ... Show Detail
- No:3490
- Date of Issue:2020/12/02 16:10
- Category: Autosampler
-
«Main Cause» A rotation error was detected in the exhaust heat fan at the top of the rear of the instrument. «Countermeasure» Press the (power switch) to tu... Show Detail
- No:3489
- Date of Issue:2020/12/02 16:10
- Category: Autosampler
-
«Main Cause» Temperature control is turned off because a rotation error was detected in the cooling fan of the autosampler. «Countermeasure» Press the (powe... Show Detail
- No:3488
- Date of Issue:2020/12/02 16:10
- Category: Autosampler
-
«Main Cause» The front panel of the autosampler may not be attached correctly. Alternatively, the sensor that detects the position of the front panel is malf... Show Detail
- No:3487
- Date of Issue:2020/12/02 16:10
- Category: Autosampler
-
«Main Cause» The consumable part count data for the autosampler and column oven is corrupted. «Countermeasure» Press the (power switch) to turn the power OF... Show Detail
- No:3486
- Date of Issue:2020/12/02 16:10
- Category: Autosampler , Column Oven
-
«Main Cause» The backup parameters for the autosampler and column oven are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail
- No:3485
- Date of Issue:2020/12/02 16:10
- Category: Autosampler , Column Oven