Frequently Referenced FAQ

FAQ within [ HPLC ]

761-770 of Total 943

≪ 77 / 95page ≫
  • Error Code:[0x907B]

    «Main Cause» The sample cooler did not stabilize at the set temperature during validation and calibration. «Countermeasure» Check whether there are any pro... Show Detail

  • Error Code:[0x907A]

    «Main Cause» There is no sample vial set at the sample vial number specified in the sequence. «Countermeasure» Set a sample vial at the sample vial number s... Show Detail

  • Error Code:[0x9079]

    «Main Cause» The column oven did not stabilize at the set temperature during validation and calibration. «Countermeasure» Check whether there are any probl... Show Detail

  • Error Code:[0x9078]

    «Main Cause» A rack number that is different from the rack number set in the sequence may have been notified to the instrument. «Countermeasure» Clear the e... Show Detail

  • Error Code:[0x9077]

    «Main Cause» A vial number that is different from the sample vial number set in the sequence may have been notified to the instrument. «Countermeasure» Clea... Show Detail

  • Error Code:[0x9076]

    «Main Cause» The injection volume setting value exceeds the [Max Injection Volume] setting in the system settings. «Countermeasure» Change the injection vol... Show Detail

  • Error Code:[0x9055]

    «Main Cause» More than 10 instruments are connected to the same group. «Countermeasure» Check the number of instruments connected to the instrument group wh... Show Detail

  • Error Code:[0x9075]

    «Main Cause» Purge was executed because the replacement pump pressure reached the [minimum pressure] in the method parameters. «Countermeasure» Inspect for ... Show Detail

  • Error Code:[0x9074]

    «Main Cause» The system check is timed out. «Countermeasure» Execute system check again. If the error is not resolved, contact a Shimadzu sales or service r... Show Detail

  • Error Code:[0x9053]

    «Main Cause» IP address renewal/refresh failed. «Countermeasure» Contact the network administrator. Set the IP address settings (automatic acquisition/manua... Show Detail