  • Error Code:[0x910C]

    «Main Cause» The backup parameters of the pump are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales o... Show Detail

  • Error Code:[0x9109]

    «Main Cause» The degassing unit is not connected. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service r... Show Detail

  • Error Code:[0x9100]

    «Main Cause» An sealing ability error occurred in the vacuum flow path of the degassing unit. «Countermeasure» Check the vacuum pressure of the degassing un... Show Detail

  • Error Code:[0x9063]

    «Main Cause» An error occurred in communication between the autosampler or column oven and controller, but the error recovered. «Countermeasure» Clear the ... Show Detail

  • Error Code:[0x9062]

    «Main Cause» An error occurred in communication between the autosampler or column oven and controller, and the communication was disconnected. «Countermeas... Show Detail

  • Error Code:[0x9061]

    «Main Cause» An error occurred in communication between the replacement pump and controller, but the error recovered. «Countermeasure» Clear the error. Show Detail

  • Error Code:[0x9060]

    «Main Cause» An error occurred in communication between the replacement pump and controller, and the communication was disconnected. «Countermeasure» Clear ... Show Detail

  • Error Code:[0x907E]

    «Main Cause» When overlapping injection is run, the sample vial is not set in the sample vial number specified for the next acquisition. «Countermeasure» C... Show Detail

  • Error Code:[0x907D]

    «Main Cause» An error occurred in communication between the detector and controller. «Countermeasure» Clear the error. If the error occurs again, contact a... Show Detail

  • Error Code:[0x907C]

    «Main Cause» The instrument power was turned OFF during acquisition. «Countermeasure» Check whether the power was turned OFF via the instrument console or t... Show Detail