Frequently Referenced FAQ
FAQ within [ HPLC ]
-
«Main Cause» An sealing ability error occurred in the vacuum flow path of the degassing unit. «Countermeasure» Check the vacuum pressure of the degassing un... Show Detail
- No:3544
- Date of Issue:2020/12/02 16:11
- Category: Solvent Delivery Module
-
«Main Cause» An error occurred in communication between the autosampler or column oven and controller, but the error recovered. «Countermeasure» Clear the ... Show Detail
- No:3543
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» An error occurred in communication between the autosampler or column oven and controller, and the communication was disconnected. «Countermeas... Show Detail
- No:3542
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» An error occurred in communication between the replacement pump and controller, but the error recovered. «Countermeasure» Clear the error. Show Detail
- No:3541
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» An error occurred in communication between the replacement pump and controller, and the communication was disconnected. «Countermeasure» Clear ... Show Detail
- No:3540
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» When overlapping injection is run, the sample vial is not set in the sample vial number specified for the next acquisition. «Countermeasure» C... Show Detail
- No:3539
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» An error occurred in communication between the detector and controller. «Countermeasure» Clear the error. If the error occurs again, contact a... Show Detail
- No:3538
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» The instrument power was turned OFF during acquisition. «Countermeasure» Check whether the power was turned OFF via the instrument console or t... Show Detail
- No:3537
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» The sample cooler did not stabilize at the set temperature during validation and calibration. «Countermeasure» Check whether there are any pro... Show Detail
- No:3536
- Date of Issue:2020/12/02 16:11
- Category: System Controller
-
«Main Cause» There is no sample vial set at the sample vial number specified in the sequence. «Countermeasure» Set a sample vial at the sample vial number s... Show Detail
- No:3535
- Date of Issue:2020/12/02 16:11
- Category: System Controller