«Main Cause» Column oven door was open for too long time while using suppressor. «Countermeasure» Close the door while using suppressor. Show Detail
- No:3556
- Date of Issue:2020/12/02 16:30
- Category: System Controller , Column Oven
-
«Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail
- No:3555
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail
- No:3554
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail
- No:3553
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail
- No:3552
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» The suppressor-related parameters are not correct. «Countermeasure» Check the values and set so that values are correct. Show Detail
- No:3551
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» Full-regeneration was operated while HPLC system was activated. «Countermeasure» Do not operate full-regeneration while HPLC system is activat... Show Detail
- No:3550
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» While full-regenerating, activate on or switch cartridge command was operated. «Countermeasure» Do not operate those commands while full regene... Show Detail
- No:3549
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» The set temperature of the column oven is set to 86ºC or higher while the column management device (CMD) is connected. «Countermeasure» Please ... Show Detail
- No:3548
- Date of Issue:2020/12/02 16:11
- Category: Column Oven
-
«Main Cause» The consumable part count data of the pump is corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu... Show Detail
- No:3547
- Date of Issue:2020/12/02 16:11
- Category: Solvent Delivery Module