«Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether PDA detector and system controller... Show Detail
- No:4094
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» Communication between the instrument and workstation was interrupted. «Countermeasure» Check the communication environment and operating status... Show Detail
- No:4095
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether a LAN cable is disconnected (PDA d... Show Detail
- No:4096
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether PDA detector and system controller... Show Detail
- No:4097
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» An instrument IP address conflict occurred with another device. «Countermeasure» Resolve the IP address conflict. If that does not resolve the ... Show Detail
- No:4098
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» The analysis start time received from the system controller is later than the current PDA detector time. «Countermeasure» Restart the PDA detec... Show Detail
- No:4099
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» The temperature-controlled cell control circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shi... Show Detail
- No:4100
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» The temperature did not decrease to the setting temperature even after five minutes of temperature control. «Countermeasure» Specify a somewhat... Show Detail
- No:4101
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» Analysis was started without connecting the flow cell. «Countermeasure» Connect the cell to the main unit. Show Detail
- No:4102
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
«Main Cause» Analysis was started without registering the cell in the main unit. «Countermeasure» Register the cell in the main unit using Utility. Show Detail
- No:4103
- Date of Issue:2020/12/02 16:45
- Category: Detector