FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Error Code:[0x5861] FATAL CELL TEMP
«Main Cause» The temperature-controlled cell control circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shi... Show Detail
- No:4054
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5863] WARNING CELL LOW SET TEMP
«Main Cause» The temperature did not decrease to the setting temperature even after five minutes of temperature control. «Countermeasure» Specify a somewhat... Show Detail
- No:4055
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5864] WARNING CELL NOT CONNECTED
«Main Cause» Analysis was started without connecting the flow cell. «Countermeasure» Connect the cell to the main unit. Show Detail
- No:4056
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5868] WARNING D2 NOT GENUINE
«Main Cause» Information in the D2 lamp memory could not be read. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail
- No:4057
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5869] WARING D2 TIME NOT UPDATED
«Main Cause» Cannot write data to D2 memory. «Countermeasure» Check the D2 lamp connection. If that does not resolve the error, replace the D2 l... Show Detail
- No:4058
- Date of Issue:2020/12/02 16:45
- Last Updated:2020/12/07 11:14
- Category: Detector
-
Error Code:[0x5870] ERROR CELL TEMP SENSOR
«Main Cause» A broken circuit may have occurred in the cell temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sale... Show Detail
- No:4059
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5878] FATAL LEAK SENSOR
«Main Cause» A leak sensor short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service represe... Show Detail
- No:4060
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5879] FATAL LEAK SENSOR
«Main Cause» A broken circuit may have occurred in the leak sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service... Show Detail
- No:4061
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5881] ERROR CELL HEATER
«Main Cause» A broken circuit may have occurred in the cell heater. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service... Show Detail
- No:4062
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5885] FATAL REAR FAN
«Main Cause» A rotation problem was detected for the fan on the back of the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shima... Show Detail
- No:4063
- Date of Issue:2020/12/02 16:45
- Category: Detector
701-710 of Total 941