Frequently Referenced FAQ

FAQ within [ HPLC ]

691-700 of Total 943

70 / 95page
  • Error Code:[0x85D1] ERR RC SLIP Z

    «Main Cause» Motor's Z-axis movement (up/down) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3606
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x85D0] ERR RC SLIP X

    «Main Cause» Motor's X-axis movement (i.e., taking out the rack) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service repre... Show Detail

    • No:3605
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8569] NDLE PROTECTED

    «Main Cause» A foreign substance was detected at the tip of the needle. «Countermeasure» Check for foreign substances inside the Autosampler. Show Detail

    • No:3604
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8566] ERR HEATER

    «Main Cause» There is an error in the sample cooler's heating unit. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representat... Show Detail

    • No:3603
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8564] ERR SLIP Y

    «Main Cause» Needle's Y-axis (forward/backward) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service represen... Show Detail

    • No:3602
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8563] ERR SLIP X

    «Main Cause» Needle's X-axis (sideways) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3601
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8551] ERR LEAK DETECT

    «Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail

    • No:3600
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8550] ERR LEAK DETECT

    «Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail

    • No:3599
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8548] STOP COOLER R

    «Main Cause» Since sample cooler R has operated without sample rack for more than 30 minutes, sample cooler R is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3598
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler
  • Error Code:[0x8547] STOP COOLER M

    «Main Cause» Since sample cooler M has operated without sample rack for more than 30 minutes, sample cooler M is stopped. «Countermeasure» Insert a sample r... Show Detail

    • No:3597
    • Date of Issue:2020/12/02 16:30
    • Category: Autosampler