Error Code:[0x85D1] ERR RC SLIP Z
«Main Cause» Motor's Z-axis movement (up/down) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3606
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85D0] ERR RC SLIP X
«Main Cause» Motor's X-axis movement (i.e., taking out the rack) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service repre... Show Detail
- No:3605
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8569] NDLE PROTECTED
«Main Cause» A foreign substance was detected at the tip of the needle. «Countermeasure» Check for foreign substances inside the Autosampler. Show Detail
- No:3604
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8566] ERR HEATER
«Main Cause» There is an error in the sample cooler's heating unit. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representat... Show Detail
- No:3603
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8564] ERR SLIP Y
«Main Cause» Needle's Y-axis (forward/backward) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service represen... Show Detail
- No:3602
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8563] ERR SLIP X
«Main Cause» Needle's X-axis (sideways) movement is incorrect. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3601
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8551] ERR LEAK DETECT
«Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail
- No:3600
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8550] ERR LEAK DETECT
«Main Cause» (1) Leak sensor was detected leakage. (2) Sensor is too sensitive. «Countermeasure» (1) Inspect the leakage from the flow line. Wipe out spille... Show Detail
- No:3599
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8548] STOP COOLER R
«Main Cause» Since sample cooler R has operated without sample rack for more than 30 minutes, sample cooler R is stopped. «Countermeasure» Insert a sample r... Show Detail
- No:3598
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8547] STOP COOLER M
«Main Cause» Since sample cooler M has operated without sample rack for more than 30 minutes, sample cooler M is stopped. «Countermeasure» Insert a sample r... Show Detail
- No:3597
- Date of Issue:2020/12/02 16:30
- Category: Autosampler