FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Error Code:[0x940B] ERROR OVER T.MAX
«Main Cause» Temperature control was stopped because the column oven temperature exceeded the upper limit temperature. «Countermeasure» Set the column oven ... Show Detail
- No:4044
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x940C] FATAL OVEN TEMP SENSOR H
«Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail
- No:4045
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x940D] FATAL OVEN TEMP SENSOR L
«Main Cause» A broken or short circuit may have occurred in the temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu ... Show Detail
- No:4046
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x940E] WARNING LOW SET TEMP
«Main Cause» The setting is too low, relative to room temperature. «Countermeasure» Check the column oven temperature setting in instrument parameter settin... Show Detail
- No:4047
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x940F] ERROR WRITE TO CMD
«Main Cause» Writing to the column management device (CMD) failed. «Countermeasure» Do not disconnect the CMD during an analysis or during writing. If a CMD... Show Detail
- No:4048
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x9410] FATAL AUX HEATER
«Main Cause» The auxiliary heater is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service repr... Show Detail
- No:4049
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x9411] ERROR DOOR SENSOR
«Main Cause» The column oven door was left ajar for a long time. «Countermeasure» Make sure the door is securely closed. If it is fully closed, it might be ... Show Detail
- No:4050
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x58E9] WARNING NOT TEMP-CONT CELL
«Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail
- No:4051
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5803] WARNING MICRO BOARD
«Main Cause» The micro cell DET board is not properly recognized. «Countermeasure» Check the cell type. To replace the current DET board with one for a micr... Show Detail
- No:4052
- Date of Issue:2020/12/02 16:45
- Category: Detector
-
Error Code:[0x5821] ERROR LEAK DETECTED
«Main Cause» There is a leak in the detector. «Countermeasure» Inspect the detector leak and wipe up any liquid from around the leak sensor. Show Detail
- No:4053
- Date of Issue:2020/12/02 16:45
- Category: Detector
691-700 of Total 941