(HPLC) Check the Condition of Consumables (System Check)
By executing the system check, you can check the usage of consumables for the entire system. • LCsolution, LabSolutions To open the [System Check... Show Detail
- No:4965
- Date of Issue:2021/02/05 10:54
- Category: Common
-
(HPLC) Non-injection method (Without sample)
LCsolution, LabSolutions, how to do non-injection when the system includes an autosampler. Vial#:Sets the vial No. of the autosampler to be used for dat... Show Detail
- No:4925
- Date of Issue:2021/02/05 10:54
- Last Updated:2021/02/09 08:52
- Category: Common
-
(HPLC) SIL Adjusting the plate position
Adjusting the plate position «ADJUST PLATE» WARNING: Do NOT put your hand inside the autosampler while the instrument is operating. During adjustm... Show Detail
- No:4816
- Date of Issue:2021/01/21 10:28
- Category: Autosampler
- Brached FAQ
-
(HPLC) SIL Setting the distance at which the needle lower...
[PARAMETER] → [NEEDLE STROKE]. or It setting by softwae. This sets the distance at which the needle lowers into a sample vial. Pressing [→] opens a set... Show Detail
- No:4825
- Date of Issue:2021/01/21 10:26
- Category: Autosampler
- Brached FAQ
-
(HPLC) Autosampler Adjusting the injection port position
Adjusting the injection port position «ADJUST INJ. P» WARINING: Do NOT put your hand inside the autosampler while the instrument is operating. Dur... Show Detail
- No:4804
- Date of Issue:2021/01/19 12:51
- Category: Autosampler
- Brached FAQ
-
1. Connect the resister tube(Example: ID0.1x4m: Parts Number: 228-32722-92)to the pump outlet instead of HPLC column. 2. Fill the reservoir with purifie... Show Detail
- No:4773
- Date of Issue:2021/01/14 10:03
- Last Updated:2021/01/16 15:43
- Category: Solvent Delivery Module
-
(HPLC) Abnormal pressure value of Solvent Delivery Module
Zero adjustment of pressure sensor. 1. Set to [ZERO ADJUST] in the [CONTROL] group. * For LC-20A / 30A, [ZERO ADJ] 2. Open the drain valve to remove ... Show Detail
- No:4768
- Date of Issue:2021/01/13 14:16
- Category: Solvent Delivery Module
-
(HPLC) SPD-M30A Handling and Cleaning of cell
<Cell Cleaning Method> 1. Flow 17 % phosphoric acid (5times dilution of 85 % phosphoric acid in water) for one hour at 1 mL/min. 2. Switch to water a... Show Detail
- No:4755
- Date of Issue:2021/01/12 14:08
- Category: Detector
-
(HPLC) RID "EN.ALRM" is displayed
• Causes: The incident light intensity to the photodiode is insufficient. * The total light intensity (TOTAL EN) is less than 5500.(In case of RID-10A, les... Show Detail
- No:4753
- Date of Issue:2021/01/12 13:56
- Category: Detector
-
(HPLC) RID "OVER" is displayed
• "OVER" is displayed in the "range" field of the initial screen. Cause: The recorder output has exceeded the scale limit. Actions: Press [zero]. ... Show Detail
- No:4750
- Date of Issue:2021/01/12 13:50
- Last Updated:2021/01/12 13:51
- Category: Detector