FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Error Code:[0x85DD] ERR RC COOLER
«Main Cause» Abnormality in cooler control unit. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3616
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85DC] ERR RC TEMP SENS
«Main Cause» There is an error in the sample cooler's temperature sensor. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representa... Show Detail
- No:3615
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85DB] ERR SET RACK
«Main Cause» Error occurred while rack was being set. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3614
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85DA] ERR CATCH RACK
«Main Cause» An error occurred while a rack is being taken out. The plate has fallen while it is being loaded or dew condensation water gathers in large quan... Show Detail
- No:3613
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85D9] ERR RELEASE RACK
«Main Cause» Rack was not released correctly. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3612
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85D6] ERR HAND HOME Y
«Main Cause» Motor's Y-axis movement (forward/backward) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3611
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85D5] ERR HAND HOME Z
«Main Cause» Motor's Z-axis movement (i.e., moving the rack vertically) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or servic... Show Detail
- No:3610
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85D4] ERR HAND HOME R
«Main Cause» Motor's R-axis movement (i.e., rotating the rack) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service represe... Show Detail
- No:3609
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85D3] ERR HAND HOME X
«Main Cause» Motor's X-axis movement (i.e., taking out the rack) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service repre... Show Detail
- No:3608
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85D2] ERR RC SLIP Y
«Main Cause» Motor's Y-axis movement (forward/backward) is incorrect. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3607
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
681-690 of Total 943