Error Code:[0x5402] FATAL AC INPUT POWER
«Main Cause» The power supply voltage or frequency was not recognized properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sale... Show Detail
- No:4024
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x5403] ERROR EXTERNAL INPUT
«Main Cause» Temperature control stopped because an error signal was input via the column oven external input terminal. «Countermeasure» Check the connected... Show Detail
- No:4025
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x5404] ERROR HIGH SET TEMP
«Main Cause» The temperature setting is too high. The temperature cannot be set higher than 85ºC if a flow channel selection valve or column management devic... Show Detail
- No:4026
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x5405] FATAL LEAK SENSOR
«Main Cause» The leak sensor cannot be detected properly. A broken or short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and ... Show Detail
- No:4027
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x5406] FATAL ROM FAILURE
«Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:4028
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x5407] FATAL RAM FAILURE
«Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:4029
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x5408] ERROR CMD CHECK SUM
«Main Cause» Data written in the column management device (CMD) is inconsistent. «Countermeasure» Disconnect or reinitialize the CMD. Show Detail
- No:4030
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x5409] FATAL ROOM TEMP SENSOR
«Main Cause» A broken or short circuit may have occurred in the room temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shim... Show Detail
- No:4031
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x540A] ERROR LEAK DETECTED
«Main Cause» Temperature control was stopped because leakage was detected in the column oven. «Countermeasure» Inspect the leakage from flow channels or oth... Show Detail
- No:4032
- Date of Issue:2020/12/02 16:45
- Category: Column Oven
-
Error Code:[0x540B] ERROR INT-VALVE2 [0x##]
«Main Cause» An error was detected in the flow channel selection valve 2 inside the column oven. (The error code "0x##" indicates a valve unit error. Refer t... Show Detail
- No:4033
- Date of Issue:2020/12/02 16:45
- Category: Column Oven