«Main Cause» Mobile phase has leaked inside the instrument. «Countermeasure» Check the plumbing. Show Detail
- No:3644
- Date of Issue:2020/12/02 16:30
- Category: Detector
-
«Main Cause» Invalid EEPROM parameter. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3643
- Date of Issue:2020/12/02 16:30
- Category: Detector
-
Error Code:[0x81F1] WARN : DEGAS PRS
«Main Cause» The abnormality of the vacuum pressure of the on-line degasser connect to this instrument will display for a certain time of time. Check the vac... Show Detail
- No:3642
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x81D3] WARN : M-PHASE
«Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail
- No:3641
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x81D2] WARN : M-PHASE
«Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail
- No:3640
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x81D1] WARN : M-PHASE
«Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail
- No:3639
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x81D0] WARN : M-PHASE
«Main Cause» The remaining volume of mobile phase is less than the alarm level. «Countermeasure» Replacement the mobile phase and reset the mobile phase qua... Show Detail
- No:3638
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x0030] ERROR P.MAX
«Main Cause» The discharge pressure of pump exceeds the Settings [P.Max]. Pumping will automatically stop. However, if "S-PROT" of auxiliary operation is Set... Show Detail
- No:3637
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x0034] ERROR LEAK
«Main Cause» Detection of fluid leak due to pump X. «Countermeasure» Inspection the leak place of flow line, etc., and wipe off the fluid around the leak se... Show Detail
- No:3636
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x0032] ERROR HOME POS
«Main Cause» Display when the home position of the pump motor cannot detection, the motor does not operate, or the motor is slipping. «Countermeasure» Once ... Show Detail
- No:3635
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
651-660 of Total 941