«Main Cause» Probable causes are the D2 lamp connector is not connected, the D2 lamp malfunctioned, or the lighting circuit malfunctioned. «Counter... Show Detail
- No:3664
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:20
- Category: Detector
-
«Main Cause» D2 lamp malfunctioned. Lighting circuit malfunctioned. «Countermeasure» Replace the D2 lamp. Click here for Replacement method o... Show Detail
- No:3663
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:19
- Category: Detector
-
«Main Cause» Lamp housing temperature sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3662
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Leak sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3661
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Leak sensor is short. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3660
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» The wavelength of the installed cell is not calibrated. «Countermeasure» Perform wavelength calibration. Show Detail
- No:3659
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» The exposure time of the installed cell is not adjusted. «Countermeasure» Set the exposure time. Show Detail
- No:3658
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» (1) An unregistered cell was installed in the instrument. (2) Failed to read cell ID. «Countermeasure» (1) Perform cell registration, exposure... Show Detail
- No:3657
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Cell is not installed or the cell cable is not connected. «Countermeasure» Install the cell and connect the cable correctly. Show Detail
- No:3656
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Temperature-controlled flow cell malfunctioned, control circuit malfunctioned. «Countermeasure» Turn the power OFF and contact a Shimadzu sales... Show Detail
- No:3655
- Date of Issue:2020/12/02 16:31
- Category: Detector