  • Error Code:[0x89D3]

    «Main Cause» (1) The suction filter on the back of the instrument is clogged. (2) Malfunction inside the instrument. «Countermeasure» (1) If the filter is ... Show Detail

    • No:3676
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D2]

    «Main Cause» PDA Detector: Temperature is increasing inside the instrument. «Countermeasure» If the filter is clogged, replace the filter. Show Detail

    • No:3675
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D1]

    «Main Cause» Failed to detect light shutter/filter origin. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3674
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89CD]

    «Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail

    • No:3673
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89CC]

    «Main Cause» The W lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. «Countermeasure» Connect the W lamp connect... Show Detail

    • No:3672
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89CB]

    «Main Cause» The D2 lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. «Countermeasure» Connect the D2 lamp conne... Show Detail

    • No:3671
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89CA]

    «Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Click here for Replacement method of Lamp (HPLC... Show Detail

    • No:3670
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:21
    • Category: Detector
  • Error Code:[0x89C9]

    «Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail

    • No:3669
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89C8]

    «Main Cause» A probable cause is that the cooling fan blast flow is poor, or the cooling fan or heater circuit malfunctioned. «Countermeasure» Ensure at le... Show Detail

    • No:3668
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89C7]

    «Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace th... Show Detail

    • No:3667
    • Date of Issue:2020/12/02 16:31
    • Category: Detector