«Main Cause» Failed to detect light shutter/filter origin. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3674
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail
- No:3673
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» The W lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. «Countermeasure» Connect the W lamp connect... Show Detail
- No:3672
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» The D2 lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. «Countermeasure» Connect the D2 lamp conne... Show Detail
- No:3671
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Click here for Replacement method of Lamp (HPLC... Show Detail
- No:3670
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:21
- Category: Detector
-
«Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail
- No:3669
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» A probable cause is that the cooling fan blast flow is poor, or the cooling fan or heater circuit malfunctioned. «Countermeasure» Ensure at le... Show Detail
- No:3668
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace th... Show Detail
- No:3667
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» The lamp cover was not in place at instrument startup. «Countermeasure» Mount the lamp cover, and turn the lamp on again. Show Detail
- No:3666
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Probable causes are the W lamp connector is not connected, or the W lamp malfunctioned. «Countermeasure» Check the W lamp connections. Check t... Show Detail
- No:3665
- Date of Issue:2020/12/02 16:31
- Category: Detector