Error Code:[0x525D] ERROR PFILE NO IF CMD
«Main Cause» There is no corresponding IF command for an ENDIF command in the pretreatment program. «Countermeasure» Add an IF command. Show Detail
- No:3974
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x525E] ERROR PFILE FOR COUNT
«Main Cause» There are more than 19 FOR commands within a FOR command in the pretreatment program. «Countermeasure» Set it to 19 or fewer. Show Detail
- No:3975
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x525F] ERROR PFILE FOR VAR
«Main Cause» The variable specified by a FOR command in the pretreatment program is incorrect. «Countermeasure» Enter the correct variable. Show Detail
- No:3976
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5260] ERROR PFILE NEXT NO FOR
«Main Cause» There is no corresponding FOR command for a NEXT command in the pretreatment program. «Countermeasure» Add a FOR command. Show Detail
- No:3977
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5261] ERROR PFILE NEXT VAR
«Main Cause» The variable specified by a FOR command in the pretreatment program does not match the variable for the NEXT command. «Countermeasure» Enter th... Show Detail
- No:3978
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5262] ERROR PFILE CALC OPE
«Main Cause» A calculation formula in the pretreatment program is missing an operand. «Countermeasure» Add an operand. Show Detail
- No:3979
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5263] ERROR PFILE CALC UNDER
«Main Cause» The result of a calculation command in the pretreatment program is a negative number. «Countermeasure» Revise the pretreatment program so that ... Show Detail
- No:3980
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5264] ERROR PFILE CALC OVER
«Main Cause» The result of a calculation command in the pretreatment program exceeded the maximum value 65536. «Countermeasure» Revise the pretreatment prog... Show Detail
- No:3981
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5265] ERROR PFILE ASSIGN RVAR
«Main Cause» The value substituted for a variable in a calculation command in the pretreatment program exceeded the maximum value 65536. «Countermeasure» Re... Show Detail
- No:3982
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5266] ERROR PFILE ASSIGN LVAR
«Main Cause» The variable assigned by an assignment statement in the pretreatment program is not correct. «Countermeasure» Enter the correct variable. Show Detail
- No:3983
- Date of Issue:2020/12/02 16:44
- Category: Autosampler