Frequently Referenced FAQ

FAQ within [ HPLC ]

611-620 of Total 943

≪ 62 / 95page ≫
  • Error Code:[0x8B07] SFE ERR DISCONNECT

    «Main Cause» The connection cable between the SFE and the rack-change is not connected properly. Or the rack-changer is turned off. «Countermeasure» Check ... Show Detail

  • Error Code:[0x8B01] SFE SYSTEM ERROR

    «Main Cause» System Error «Countermeasure» Press the (power button) on the instrument to turn the power OFF and contact a Shimadzu sales or service represe... Show Detail

  • Error Code:[0x8F02]

    «Main Cause» Liquid leakage has been detected. «Countermeasure» (1) Remove the plastic tray or metal tray, and wipe the liquid accumulated on the leakage tr... Show Detail

  • Error Code:[0x8F01]

    «Main Cause» No empty vial in the fraction collector. «Countermeasure» (1) Fraction collector has used up to the final vial. Check the final vial (F.VIAL) s... Show Detail

  • Error Code:[0x89DA]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3682
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D9]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3681
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D8]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3680
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D7]

    «Main Cause» Power heat exhaust fan malfunction. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3679
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D6]

    «Main Cause» Lamp compartment fan malfunction «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3678
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D4]

    «Main Cause» (1)D2 lamp malfunction. (2)Malfunction inside the instrument. «Countermeasure» (1) Replace the D2 lamp. Click here for Replace... Show Detail

    • No:3677
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:22
    • Category: Detector