FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Error Code:[0x5235] ERROR DETECT CLOGGING
«Main Cause» A blockage was detected in the high pressure flow channels inside the autosampler. «Countermeasure» Perform auto-purge. Show Detail
- No:3934
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5236] ERROR MISMATCH PLATE
«Main Cause» The sample plate type does not match the type specified in the method file or the barcode label attached to the sample plate does not match the ... Show Detail
- No:3935
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5237] ERROR PFILE ILLEGAL CMD
«Main Cause» A command not present in the pretreatment program is specified. «Countermeasure» Check the command. Show Detail
- No:3936
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5238] ERROR PFILE ILLEGAL VAL
«Main Cause» A variable not present in the pretreatment program is specified. «Countermeasure» Check the variable. Show Detail
- No:3937
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5239] ERROR PFILE PURGE VOL
«Main Cause» The liquid volume specified by the PURGE.RP or PURGE.ML command in the pretreatment program exceeds the setting range. «Countermeasure» Change ... Show Detail
- No:3938
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x523A] ERROR PFILE PURGE SPEED
«Main Cause» The suction speed specified by the PURGE.RP or PURGE.ML command in the pretreatment program exceeds the setting range. «Countermeasure» Change ... Show Detail
- No:3939
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x523B] ERROR PFILE ASPIR VOL
«Main Cause» The liquid volume specified for suction in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value w... Show Detail
- No:3940
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x523C] ERROR PFILE ASPIR SPEED
«Main Cause» The suction speed specified for suction in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value w... Show Detail
- No:3941
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x523D] ERROR PFILE DISP VOLUME
«Main Cause» The liquid volume specified for discharge in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value... Show Detail
- No:3942
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x523E] ERROR PFILE DISP SPEED
«Main Cause» The discharge speed specified for discharge in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a val... Show Detail
- No:3943
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
581-590 of Total 943