Error Code:[0x522B] WARNING RINSE SOLUTION R1
«Main Cause» The remaining level of rinse solution R1 is low. «Countermeasure» Add more rinse solution R1. Show Detail
- No:3924
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x522C] WARNING RINSE SOLUTION R2
«Main Cause» The remaining level of rinse solution R2 is low. «Countermeasure» Add more rinse solution R2. Show Detail
- No:3925
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x522D] WARNING RINSE SOLUTION R3
«Main Cause» The remaining level of rinse solution R3 is low. «Countermeasure» Add more rinse solution R3. Show Detail
- No:3926
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x522E] WARNING PLATE ADJUST
«Main Cause» The needle position was not adjusted based on the given sample plate. «Countermeasure» Adjust the needle position based on the given sample plate. Show Detail
- No:3927
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x522F] WARNING PLATE ID
«Main Cause» The type of sample plate placed on the sample rack was not read correctly. «Countermeasure» Reposition the sample plate. Show Detail
- No:3928
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5230] WARNING SYRINGE
«Main Cause» Syringe use was specified in autosampler settings, but no syringe is connected. «Countermeasure» Connect a syringe to the autosampler and switc... Show Detail
- No:3929
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5231] ERROR VIAL NUMVER
«Main Cause» The specified vial number exceeds the number of vials that can be placed on the sample plate. «Countermeasure» Change the vial number setting. Show Detail
- No:3930
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5232] ERROR INJ VOLUME
«Main Cause» The injection volume exceeds the maximum injection volume setting value. «Countermeasure» Change the injection volume to a value less than the ... Show Detail
- No:3931
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5233] ERROR PFILE
«Main Cause» The pretreatment program size exceeded the limit value (max. 255 bytes per file or 4,096 bytes for the entire program). «Countermeasure» Optimi... Show Detail
- No:3932
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5234] WARNING DETECT CLOGGING
«Main Cause» A blockage was detected in the high pressure flow channels inside the autosampler. «Countermeasure» Perform auto-purge. Show Detail
- No:3933
- Date of Issue:2020/12/02 16:44
- Category: Autosampler