Frequently Referenced FAQ

FAQ within [ HPLC ]

561-570 of Total 943

≪ 57 / 95page ≫
  • Error Code:[0x00A1] ERR LEAK DETECTED

    «Main Cause» Leak detected. «Countermeasure» Inspect and repair plumbing. Wipe away leakage. Show Detail

    • No:3736
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x00A0] ERR ZG HOME POS

    «Main Cause» The home position of the zero glass drive unit cannot be properly detected. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3735
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x008C]

    «Main Cause» SUBC: Leak sensor was detected leakage. «Countermeasure» Inspect the leakage from the flow line. Wipe out spilled liquid after fixing the leak... Show Detail

  • Error Code:[0x0086]

    «Main Cause» The setting for suction/discharge of Syringe Unit is out of the range. «Countermeasure» Check the setting for suction/discharge of Syringe Unit. Show Detail

  • Error Code:[0x0085]

    «Main Cause» No Motor Home Position Error of Syringe unit has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn ... Show Detail

  • Error Code:[0x0084]

    «Main Cause» Valve home position error has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the power OFF and c... Show Detail

  • Error Code:[0x0083]

    «Main Cause» No Motor Home Position Error of RV.D has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

  • Error Code:[0x0082]

    «Main Cause» No Motor Home Position Error of RV.C has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

  • Error Code:[0x0081]

    «Main Cause» No Motor Home Position Error of RV.B has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

  • Error Code:[0x0080]

    «Main Cause» No Motor Home Position Error of RV.A has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

561-570 of Total 943