Error Code:[0x870E] ERR EEPROM WRITE
«Main Cause» EEPROM write error is occurred. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3744
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x00C5] ERR BG2 TOO HIGH
«Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail
- No:3743
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x00C4] ERR BG1 TOO HIGH
«Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail
- No:3742
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x00C3] ERR SUP2 HI VOLT
«Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail
- No:3741
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/04 09:39
- Category: Detector
-
Error Code:[0x00C2] ERR SUP1 HI VOLT
«Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail
- No:3740
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/04 09:35
- Category: Detector
-
Error Code:[0x00C1] ERR VALVE2 HOME
«Main Cause» The home position of the suppressor Ch2 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail
- No:3739
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x00C0] ERR VALVE1 HOME
«Main Cause» The home position of the suppressor Ch1 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail
- No:3738
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x00A4] ERR OVER HEAT (0)
«Main Cause» Electric circuit was damaged. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3737
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x00A1] ERR LEAK DETECTED
«Main Cause» Leak detected. «Countermeasure» Inspect and repair plumbing. Wipe away leakage. Show Detail
- No:3736
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x00A0] ERR ZG HOME POS
«Main Cause» The home position of the zero glass drive unit cannot be properly detected. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3735
- Date of Issue:2020/12/02 16:31
- Category: Detector