• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ HPLC ]

551-560 of Total 943

56 / 95page
  • Error Code:[0x8714] ERR Xe LAMP

    «Main Cause» Displayed when the Xe lamp cannot be turn on. «Countermeasure» Turn the power OFF and back ON, then turn the Xe lamp on. The Xe lamp cannot be ... Show Detail

    • No:3746
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8710] ERR LM HOME

    «Main Cause» Displayed when the home position sensor of the lamp mirror is not operating correctly. «Countermeasure» Turn the power OFF and then restart the... Show Detail

    • No:3745
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x870E] ERR EEPROM WRITE

    «Main Cause» EEPROM write error is occurred. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3744
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x00C5] ERR BG2 TOO HIGH

    «Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail

    • No:3743
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x00C4] ERR BG1 TOO HIGH

    «Main Cause» GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if backgroun... Show Detail

    • No:3742
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x00C3] ERR SUP2 HI VOLT

    «Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail

    • No:3741
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/04 09:39
    • Category: Detector
  • Error Code:[0x00C2] ERR SUP1 HI VOLT

    «Main Cause» The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. ... Show Detail

    • No:3740
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/04 09:35
    • Category: Detector
  • Error Code:[0x00C1] ERR VALVE2 HOME

    «Main Cause» The home position of the suppressor Ch2 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail

    • No:3739
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x00C0] ERR VALVE1 HOME

    «Main Cause» The home position of the suppressor Ch1 valve failed to detection. «Countermeasure» (1) Disassemble the valve and clean or replace the rotor an... Show Detail

    • No:3738
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x00A4] ERR OVER HEAT (0)

    «Main Cause» Electric circuit was damaged. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail

    • No:3737
    • Date of Issue:2020/12/02 16:31
    • Category: Detector

551-560 of Total 943

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.