Error Code:[0x5217] ERROR TEMP SENSOR2
«Main Cause» Temperature sensor 2 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail
- No:3904
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5218] FATAL PELTIER1 VOLTAGE
«Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail
- No:3905
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5219] FATAL PELTIER2 VOLTAGE
«Main Cause» The Peltier element used for temperature adjustment is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a... Show Detail
- No:3906
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x521A] FATAL INNER FAN
«Main Cause» A fan rotation problem was detected for the fan inside the cabinet. «Countermeasure» Check for foreign matter caught in the fan in the cabinet.... Show Detail
- No:3907
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x521B] FATAL OUTER FAN
«Main Cause» A rotation problem was detected for the fan on the back of the instrument. «Countermeasure» Check for foreign matter caught in the fan on the b... Show Detail
- No:3908
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x521C] FATAL POWER FAN
«Main Cause» A fan rotation problem was detected in the power supply fan. «Countermeasure» Check for foreign matter caught in the power supply fan. If that ... Show Detail
- No:3909
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x521D] WARNING PELTIER STOP
«Main Cause» No sample rack was detected for a given period. «Countermeasure» Place the sample rack and start temperature control. Show Detail
- No:3910
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x521E] ERROR LEAK DETECTED
«Main Cause» Leakage occurred from a tubing joint or valve gap. «Countermeasure» Check the following locations for leakage. ・If there is leakage from the in... Show Detail
- No:3911
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x521F] FATAL LEAK SENSOR
«Main Cause» The leak sensor cannot be detected properly. A broken circuit may have occurred in the leak sensor. «Countermeasure» Switch the instrument powe... Show Detail
- No:3912
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5220] FATAL LEAK SENSOR
«Main Cause» The leak sensor cannot be detected properly. The leak sensor might not be connected. «Countermeasure» Switch the instrument power OFF and conta... Show Detail
- No:3913
- Date of Issue:2020/12/02 16:44
- Category: Autosampler