Error Code:[0x520D] ERROR HPV TYPE
«Main Cause» The type of high pressure valve is not correct. «Countermeasure» Install a 6-port 2-position valve. Show Detail
- No:3894
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x520E] ERROR LPV HOME POSITION
«Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail
- No:3895
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x520F] ERROR LPV SLIP
«Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail
- No:3896
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5210] ERROR LPV DATA
«Main Cause» The low pressure valve is disconnected or there is an electrical contact problem. «Countermeasure» Check whether the low pressure valve is disc... Show Detail
- No:3897
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5211] ERROR LPV TYPE
«Main Cause» The type of low pressure valve is incorrect. «Countermeasure» Install a 7-port valve. Show Detail
- No:3898
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5212] ERROR PUMP HOME POS
«Main Cause» Plunger wear has increased or there is a problem in the drive unit. «Countermeasure» Click the [CE] button and try again. If the error is not r... Show Detail
- No:3899
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5213] ERROR ADJUST PUMP
«Main Cause» Plunger wear has increased or there is a problem in the drive unit. «Countermeasure» Click the [CE] button and try again. If the error is not r... Show Detail
- No:3900
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5214] ERROR PUMP SLIP
«Main Cause» Plunger wear has increased or there is a problem in the drive unit. «Countermeasure» Click the [CE] button and try again. If the error is not r... Show Detail
- No:3901
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5215] FATAL NEEDLE PROTECTED
«Main Cause» The needle cannot be lowered because an obstruction other than the sample vial or injection port was detected. «Countermeasure» Switch the inst... Show Detail
- No:3902
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5216] ERROR TEMP SENSOR1
«Main Cause» Temperature sensor 1 inside the cabinet detected a temperature outside the specified range. The temperature sensor malfunctioned or the temperat... Show Detail
- No:3903
- Date of Issue:2020/12/02 16:44
- Category: Autosampler