Frequently Referenced FAQ

FAQ within [ HPLC ]

491-500 of Total 943

≪ 50 / 95page ≫
  • Error Code:[0x8010]

    «Main Cause» The pump was purged because pressure did not reach the lower pressure limit value, but an error occurred again, so solvent delivery was stopped.... Show Detail

  • Error Code:[0x8003]

    «Main Cause» Solvent delivery was stopped because the pressure exceeded the system upper pressure limit value. «Countermeasure» Change the syste... Show Detail

  • Error Code:[0x8002]

    «Main Cause» The date and time specified in the startup has passed the present date and time. «Countermeasure» Specify a future date/time as the startup dat... Show Detail

  • Error Code:[0x8001]

    «Main Cause» There was no response from a instrument for a given time during a system check. «Countermeasure» Check the instrument connection status and per... Show Detail

  • Error Code:[0x4004]

    «Main Cause» Because of communication error, communication with the workstation could not be established. «Countermeasure» Check the communication environme... Show Detail

  • Error Code:[0x4001]

    «Main Cause» Because of communication error, communication with the workstation could not be established. «Countermeasure» Check the communication environme... Show Detail

  • Error Code:[0x2F37]

    «Main Cause» A communication error occurred with the instrument. «Countermeasure» Check the optical link cable. If that does resolve the error, switch OFF t... Show Detail

  • Error Code:[0x2F36]

    «Main Cause» A communication error occurred with the instrument. «Countermeasure» Check the optical link cable. If that does resolve the error, switch OFF t... Show Detail

  • Error Code:[0x2F35]

    «Main Cause» A communication error occurred with the instrument. «Countermeasure» Check the optical link cable. If that does resolve the error, switch OFF t... Show Detail

  • Error Code:[0x2F34]

    «Main Cause» A communication error occurred with the instrument. «Countermeasure» Check the optical link cable. If that does resolve the error, switch OFF t... Show Detail

491-500 of Total 943