«Main Cause» The remaining volume of rinse solution is less than the check level. «Countermeasure» Replace the rinse solution or reset the rinse solution vo... Show Detail
- No:3814
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Autosampler
-
«Main Cause» The PDA detector independently started up the system controller during an analysis. «Countermeasure» Either stop the PDA detector analysis or w... Show Detail
- No:3813
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Detector
-
«Main Cause» The PDA detector system name was changed during use. «Countermeasure» The parent-child relationship might have failed between the system contro... Show Detail
- No:3812
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Detector
-
«Main Cause» The system name setting for the master PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed be... Show Detail
- No:3811
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Detector
-
«Main Cause» The IP address for the PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed between the system... Show Detail
- No:3810
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Detector
-
«Main Cause» Multiple PDA detectors attempted to connect to the same system controller (redundant parent-child relationships). «Countermeasure» Change PDA d... Show Detail
- No:3809
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Detector
-
«Main Cause» Eleven or more system controllers and PDA detectors are assigned to the same group. «Countermeasure» Specify ten or fewer devices per group. Show Detail
- No:3808
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Detector
-
«Main Cause» A separate workstation than connected to the system controller is connected to the PDA detector. (System controllers and PDA detectors with an e... Show Detail
- No:3807
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Detector
-
«Main Cause» The pump was purged because pressure did not reach the lower pressure limit value, but an error occurred again, so solvent delivery was stopped.... Show Detail
- No:3806
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Solvent Delivery Module
-
«Main Cause» Solvent delivery was stopped because the pressure exceeded the system upper pressure limit value. «Countermeasure» Change the syste... Show Detail
- No:3805
- Date of Issue:2020/12/02 16:43
- Last Updated:2020/12/04 18:40
- Category: System Controller , Solvent Delivery Module