«Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail
- No:3824
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Autosampler
-
«Main Cause» The autosampler was unable to recognize the rack or the rack is not located in the autosampler. «Countermeasure» Check the rack. Show Detail
- No:3823
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Autosampler
-
«Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail
- No:3822
- Date of Issue:2020/12/02 16:43
- Category: System Controller , Autosampler
-
«Main Cause» The system controller attempted to start an analysis without an instrument specified in [link check] settings being connected. «Countermeasure»... Show Detail
- No:3821
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» An error signal was input from an external device connected to the system controller external input/output terminal ([IN] terminal). «Counterme... Show Detail
- No:3820
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail
- No:3819
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail
- No:3818
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail
- No:3817
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail
- No:3816
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» The backup battery for the system controller is running low. «Countermeasure» Replace the battery early, because parameter settings cannot be s... Show Detail
- No:3815
- Date of Issue:2020/12/02 16:43
- Category: System Controller