Frequently Referenced FAQ

FAQ within [ HPLC ]

471-480 of Total 943

≪ 48 / 95page ≫
  • Error Code:[0x4101]

    «Main Cause» (Only if used in the DHCP mode.) An IP address was not assigned by the DHCP server during startup. «Countermeasure» (1) Check the communication... Show Detail

  • Error Code:[0x0026]

    «Main Cause» The fraction collector starting vial number ([I.VIAL]) or ending vial number ([F.VIAL]) setting is out of range for the given rack. «Countermea... Show Detail

  • Error Code:[0x001D]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

  • Error Code:[0x0019]

    «Main Cause» The autosampler was unable to recognize the rack or the rack is not located in the autosampler. «Countermeasure» Check the rack. Show Detail

  • Error Code:[0x000F]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

  • Error Code:[0x000B]

    «Main Cause» The system controller attempted to start an analysis without an instrument specified in [link check] settings being connected. «Countermeasure»... Show Detail

  • Error Code:[0x0008]

    «Main Cause» An error signal was input from an external device connected to the system controller external input/output terminal ([IN] terminal). «Counterme... Show Detail

  • Error Code:[0x0006]

    «Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail

  • Error Code:[0x0005]

    «Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail

  • Error Code:[0x0004]

    «Main Cause» The method file, sequence file, and configuration settings were reinitialized (cleared). «Countermeasure» Check or reset method file, sequence ... Show Detail