  • Error Code:[0x001D]

    «Main Cause» A vial number specified in the sequence is out of range. «Countermeasure» Specify vial numbers that correspond to the given rack. Show Detail

  • Error Code:[0x0026]

    «Main Cause» The fraction collector starting vial number ([I.VIAL]) or ending vial number ([F.VIAL]) setting is out of range for the given rack. «Countermea... Show Detail

  • Error Code:[0x4101]

    «Main Cause» (Only if used in the DHCP mode.) An IP address was not assigned by the DHCP server during startup. «Countermeasure» (1) Check the communication... Show Detail

  • Error Code:[0x4102]

    «Main Cause» (Only if used in the DHCP mode.) An IP address was not assigned by the DHCP server when the IP address lease was extended. «Countermeasure» To ... Show Detail

  • Error Code:[0x4103]

    «Main Cause» (Only if used in the DHCP mode.) An IP address was not assigned by the DHCP server when the IP address lease was extended. «Countermeasure» To ... Show Detail

  • Error Code:[0x80F1]

    «Main Cause» The system controller cannot connect to the PDA detector due to a communication problem. «Countermeasure» (1) Check the communication environme... Show Detail

  • Error Code:[0x80F2]

    «Main Cause» The system controller cannot connect to the PDA detector due to a communication problem. «Countermeasure» (1) Check the communication environme... Show Detail

  • Error Code:[0x856A]

    «Main Cause» The injection volume exceeded the [injection volume upper limit] setting value. «Countermeasure» Inject a volume that does not exceed the [inje... Show Detail

  • Error Code:[0x2E01]

    «Main Cause» An undefined error occurred. «Countermeasure» Inspect the instrument where the error occurred and eliminate the cause of the error. Show Detail

  • Error Code:[0x2E02]

    «Main Cause» An undefined error occurred. «Countermeasure» Inspect the instrument where the error occurred and eliminate the cause of the error. Show Detail