Error Code:[0x5000] FATAL CONNECTION A-B
«Main Cause» Communication failed between pumps A and B in the main system. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or... Show Detail
- No:3846
- Date of Issue:2020/12/02 16:43
- Category: Solvent Delivery Module
-
«Main Cause» The zero point value has deviated significantly from the reference value. «Countermeasure» Please ensure there are no objects placed on the bot... Show Detail
- No:3845
- Date of Issue:2020/12/02 16:43
- Category: Mobile Phase Monitor
-
«Main Cause» The measurement value is out of range. «Countermeasure» Check that the bottle holder is properly connected and then unplug and reconnect the po... Show Detail
- No:3844
- Date of Issue:2020/12/02 16:43
- Category: Mobile Phase Monitor
-
«Main Cause» The system was started without a previously connected bottle holder connected. Check that the sensors to be used are properly connected. «Count... Show Detail
- No:3843
- Date of Issue:2020/12/02 16:43
- Category: Mobile Phase Monitor
-
«Main Cause» A communication error occurred with the mobile phase monitor. «Countermeasure» Check that the bottle holder is properly connected and then unpl... Show Detail
- No:3842
- Date of Issue:2020/12/02 16:43
- Category: Mobile Phase Monitor
-
«Main Cause» The remaining mobile phase volume is below the error level. «Countermeasure» Replace the mobile phase. Show Detail
- No:3841
- Date of Issue:2020/12/02 16:43
- Category: Mobile Phase Monitor
-
«Main Cause» The remaining mobile phase volume is below the warning level. «Countermeasure» Replace the mobile phase. Show Detail
- No:3840
- Date of Issue:2020/12/02 16:43
- Category: Mobile Phase Monitor
-
«Main Cause» An A/D board failure was detected. «Countermeasure» Switch the system controller main power supply OFF and contact a Shimadzu sales or service ... Show Detail
- No:3839
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» An A/D board failure was detected. «Countermeasure» Switch the system controller main power supply OFF and contact a Shimadzu sales or service ... Show Detail
- No:3838
- Date of Issue:2020/12/02 16:43
- Category: System Controller
-
«Main Cause» An A/D board failure was detected. «Countermeasure» Switch the system controller main power supply OFF and contact a Shimadzu sales or service ... Show Detail
- No:3837
- Date of Issue:2020/12/02 16:43
- Category: System Controller