Frequently Referenced FAQ

FAQ within [ HPLC ]

411-420 of Total 943

≪ 42 / 95page ≫
  • Error Code:[0x874B] ERR OVER HEAT

    «Main Cause» Instrument interior temperature has risen to an abnormal level. «Countermeasure» Check to be sure that the rear fan can move, and that the exh... Show Detail

    • No:3764
    • Date of Issue:2020/12/02 16:32
    • Category: Detector
  • Error Code:[0x87D0] LOW SET TEMP

    «Main Cause» This warning is displayed when the flow cell temperature fails to coincide with temperature settings after undergoing 5 minutes or more of tempe... Show Detail

    • No:3765
    • Date of Issue:2020/12/02 16:32
    • Category: Detector
  • Error Code:[0x0070] ERR OVER T.MAX

    «Main Cause» Temperature control is turned off because the temperature of the column oven exceeded the maximum temperature. «Countermeasure» Settings the in... Show Detail

    • No:3766
    • Date of Issue:2020/12/02 16:32
    • Category: Column Oven
  • Error Code:[0x000A]

    «Main Cause» Connection to a instrument with the [link check] checkbox selected wa disconnected. «Countermeasure» Please check the optical link cable. In pa... Show Detail

  • Error Code:[0x2F01]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

  • Error Code:[0x2F03]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

  • Error Code:[0x2F05]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

  • Error Code:[0x2F06]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

  • Error Code:[0x2F08]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail

  • Error Code:[0x2F09]

    «Main Cause» A communication error occurred with instrument. «Countermeasure» Check the optical link cable, turn off the main power of the system controller... Show Detail