Error Code:[0x520D] ERROR HPV TYPE
«Main Cause» The type of high pressure valve is not correct. «Countermeasure» Install a 6-port 2-position valve. Show Detail
- No:3894
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x520C] ERROR HPV DATA
«Main Cause» The high pressure valve is disconnected or there is an electrical contact problem. «Countermeasure» Check whether the high pressure valve is di... Show Detail
- No:3893
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x520B] ERROR HPV SLIP
«Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail
- No:3892
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x520A] ERROR HPV HOME POSITION
«Main Cause» The torque required to rotate the valve increased or there is a mechanical problem in the drive unit. «Countermeasure» Click the [CE] button an... Show Detail
- No:3891
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5209] FATAL Z HOME POSITION
«Main Cause» Something is obstructing the needle's line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Check for any obstr... Show Detail
- No:3890
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5208] ERROR Y SLIP
«Main Cause» Something is obstructing the Y-direction (front-rear) line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Che... Show Detail
- No:3889
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5207] ERROR Y HOME POSITION
«Main Cause» Something is obstructing the needle's line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Check for any obstr... Show Detail
- No:3888
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5206] ERROR X SLIP
«Main Cause» Something is obstructing the X-direction (left-right) line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Che... Show Detail
- No:3887
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5205] ERROR X HOME POSITION
«Main Cause» Something is obstructing the needle's line of movement or there is a mechanical problem in the drive unit. «Countermeasure» Check for any obstr... Show Detail
- No:3886
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5204] FATAL NOT CONNECT SENS
«Main Cause» Communication with the plate reading board failed. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail
- No:3885
- Date of Issue:2020/12/02 16:44
- Category: Autosampler