Frequently Referenced FAQ

FAQ within [ HPLC ]

401-410 of Total 943

≪ 41 / 95page ≫
  • Error Code:[0x8720] ERR FAN STOP

    «Main Cause» Displayed when the cooling fan for the Xe lamp's power supply stops. «Countermeasure» Check if the cooling fan is working . If the exhaust vent... Show Detail

    • No:3754
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8721] LOW SET TEMP

    «Main Cause» Displayed if the room temperature exceeds the set temperature for the flow cell by 10ºC or more. «Countermeasure» Set a flow cell temperature t... Show Detail

    • No:3755
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8723] ERR COOLER

    «Main Cause» Displayed when the temperature of the flow cell has failed to reach the set temperature within 30 minutes after the power was turned ON. «Count... Show Detail

    • No:3756
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8724] ERR HEATER

    «Main Cause» Displayed when the temperature of the flow cell has failed to reach the set temperature within 30 minutes after the power was turned ON. «Count... Show Detail

    • No:3757
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8725] ERR LEAK SENS

    «Main Cause» The leak sensor has malfunctioned, probably due to a broken/disconnected wire. «Countermeasure» Turn the power OFF and then restart the instrum... Show Detail

    • No:3758
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8726] ERR LEAK SENS2

    «Main Cause» The leak sensor has malfunctioned, probably due to a short circuit. «Countermeasure» Turn the power OFF and then restart the instrument. If the... Show Detail

    • No:3759
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8728] ERR TEMP SENS

    «Main Cause» The temperature sensor in the detector has failed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3760
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8729] ERR No.****

    «Main Cause» The PCB or other circuit of the detector has failed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3761
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8743] ERR D2 LAMP

    «Main Cause» Fault in deuterium lamp or in its circuits. «Countermeasure» Click here for Replacement method of Lamp (HPLC) Turn the power OFF... Show Detail

    • No:3762
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:24
    • Category: Detector
  • Error Code:[0x8744] ERR OVER CURRENT

    «Main Cause» Abnormally high current flowing through lamp. «Countermeasure» Click here for Replacement method of Lamp (HPLC) Turn the power O... Show Detail

    • No:3763
    • Date of Issue:2020/12/02 16:32
    • Last Updated:2020/12/07 11:24
    • Category: Detector