  • Error Code:[0x004B] ERR OVER HEAT

    «Main Cause» The temperature of the measuring point rises to 63ºC or more. «Countermeasure» Check the operating environment. If you cannot find any problem... Show Detail

    • No:3724
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x004C] ERR LAMP

    «Main Cause» Fault in Xe lamp or in its circuits. «Countermeasure» Check the Xe lamp operating time and replace it. Click here for Replacemen... Show Detail

    • No:3725
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:23
    • Category: Detector
  • Error Code:[0x0050] DATA NOT EXIST

    «Main Cause» Spectrum data was requested for output from a file which dose not contain scanned data. «Countermeasure» Check the file number set when scanni... Show Detail

    • No:3726
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x0080]

    «Main Cause» No Motor Home Position Error of RV.A has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

  • Error Code:[0x0081]

    «Main Cause» No Motor Home Position Error of RV.B has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

  • Error Code:[0x0082]

    «Main Cause» No Motor Home Position Error of RV.C has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

  • Error Code:[0x0083]

    «Main Cause» No Motor Home Position Error of RV.D has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the powe... Show Detail

  • Error Code:[0x0084]

    «Main Cause» Valve home position error has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn the power OFF and c... Show Detail

  • Error Code:[0x0085]

    «Main Cause» No Motor Home Position Error of Syringe unit has occurred. «Countermeasure» Turn the power OFF and then ON again. If the error persists, turn ... Show Detail

  • Error Code:[0x0086]

    «Main Cause» The setting for suction/discharge of Syringe Unit is out of the range. «Countermeasure» Check the setting for suction/discharge of Syringe Unit. Show Detail