Error Code:[0x8BB0] SFC ERR LEAK DETECT
«Main Cause» Eluent leak can be detected in the BPR. «Countermeasure» Stop the pump flow and then execute any action to stop the eluent leak. See detail In... Show Detail
- No:3714
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8BB1] SFC ERR LEAK SENS
«Main Cause» The leak sensor detection value is abnormal. The leak sensor may be disconnected. «Countermeasure» Press the (power button) on the instrument ... Show Detail
- No:3715
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x0040] ERR GR HOME POS
«Main Cause» Motor home position sensor does not operate properly. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3716
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x0041] ERR FIL HOME POS
«Main Cause» Filter home position sensor does not operate properly. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3717
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x0042] ERR LAMP HOME PO
«Main Cause» Lamp home position sensor does not operate properly. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3718
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x0043] ERR D2 LAMP
«Main Cause» Fault in deuterium lamp or in its circuits. «Countermeasure» Check the total lamp illumination time, and replace the lamp. Click... Show Detail
- No:3719
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:22
- Category: Detector
-
Error Code:[0x0044] ERR OVER CURRENT
«Main Cause» Abnormally high current flowing through lamp. «Countermeasure» Check the total lamp illumination time, and replace the lamp. Cli... Show Detail
- No:3720
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:23
- Category: Detector
-
Error Code:[0x0047] ERR EX HOME POS
«Main Cause» Ex grating motor has slipped. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3721
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x0048] ERR EM HOME POS
«Main Cause» Em grating motor has slipped. «Countermeasure» Contact a Shimadzu sales or service representative Show Detail
- No:3722
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
Error Code:[0x004A] CHECK NO GOOD
«Main Cause» Wavelength discrepancy exceeds 1nm. «Countermeasure» 1. Cell is not installed properly - Install cell properly 2. Error in wavelength calibrat... Show Detail
- No:3723
- Date of Issue:2020/12/02 16:31
- Category: Detector