Error Code:[0x8B40] SFE NO VESSEL
«Main Cause» No extraction vessel was placed in the rack position. «Countermeasure» Set the extraction vessel properry. If the error message is displayed a... Show Detail
- No:3694
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B41] SFE ERR T.MAX H*
«Main Cause» The temperature of the extraction vessel exceeded the [Maximum Temperature] of the instrument parameters. «Countermeasure» Maximum temperature... Show Detail
- No:3695
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B42] SFE ERR TEMP SENS H*
«Main Cause» The extraction vessel temperature sensor detection value is abnormal. The CO2 gas sensor may be disconnected. «Countermeasure» Press the (powe... Show Detail
- No:3696
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B43] SFE ERR T.MAX TUBE
«Main Cause» The temperature of the pre-heater exceeded the [Maximum Temperature] of the instrument parameters. «Countermeasure» Maximum temperature must b... Show Detail
- No:3697
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B44] SFE ERR TEMP SENS TB
«Main Cause» The pre-heater temperature sensor detection value is abnormal. The pre-heater temperature sensor may be disconnected. «Countermeasure» Press th... Show Detail
- No:3698
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B45] SFE LOW SET TEMP H*
«Main Cause» Set temperature value is inappropriate. Temperature is specified with lower than the room temperature plus 5ºC. «Countermeasure» Set temperatur... Show Detail
- No:3699
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B46] SFE ERR LEAK VESSEL*
«Main Cause» The temperature of the extraction container has decreased. Check if the extraction pack is tightened. «Countermeasure» Please confirm the extra... Show Detail
- No:3700
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B47] SFE ERR HEATER H*
«Main Cause» The extraction pack cell temperature control heater has failed. «Countermeasure» Press the (power button) on the instrument to turn the power O... Show Detail
- No:3701
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B48] SFE ERR HEATER TUBE
«Main Cause» The preheater is abnormal. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and contact a Shimadzu sales or se... Show Detail
- No:3702
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph
-
Error Code:[0x8B49] SFE ERR TEMP SENS VE
«Main Cause» The vessel thermosensor is abnormal. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and contact a Shimadzu s... Show Detail
- No:3703
- Date of Issue:2020/12/02 16:31
- Category: Supercritical Fluid Chromatograph