Error Code:[0x525D] ERROR PFILE NO IF CMD
«Main Cause» There is no corresponding IF command for an ENDIF command in the pretreatment program. «Countermeasure» Add an IF command. Show Detail
- No:3974
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x525C] ERROR PFILE TOO MANY IF
«Main Cause» The pretreatment program has more than 19 IF commands in its IF command block. «Countermeasure» Set it to 19 or fewer. Show Detail
- No:3973
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x525B] ERROR PFILE 2nd S.INJ
«Main Cause» Analysis was started two or more times in the pretreatment program. «Countermeasure» Revise the pretreatment program so that the analysis start... Show Detail
- No:3972
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x525A] ERROR PFILE EVENT VALUE
«Main Cause» The setting value for the EVENT command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value w... Show Detail
- No:3971
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5259] ERROR PFILE CALL TIMES
«Main Cause» The CALL command calls more than 19 times in the pretreatment program. «Countermeasure» Set it to 19 or fewer times. Show Detail
- No:3970
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5258] ERROR PFILE CALL FILE #
«Main Cause» The file number specified by the CALL command in the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a v... Show Detail
- No:3969
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5257] ERROR PFILE WAIT.RT RNG
«Main Cause» The wait time specified by the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value within the specif... Show Detail
- No:3968
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5256] ERROR PFILE WAIT.RT RUN
«Main Cause» An analysis time was specified before the analysis is started by the pretreatment program. «Countermeasure» Revise the pretreatment program so ... Show Detail
- No:3967
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5255] ERROR PFILE WAIT.SEC
«Main Cause» The wait time specified by the pretreatment program exceeds the setting range. «Countermeasure» Change the setting to a value within the specif... Show Detail
- No:3966
- Date of Issue:2020/12/02 16:44
- Category: Autosampler
-
Error Code:[0x5254] ERROR PFILE GOTO FOR.C
«Main Cause» The FOR command line number was specified by the GOTO command in the FOR-NEXT range in the pretreatment program. «Countermeasure» Revise the li... Show Detail
- No:3965
- Date of Issue:2020/12/02 16:44
- Category: Autosampler