FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
«Main Cause» Failed to detect light shutter/filter origin. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3674
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» PDA Detector: Temperature is increasing inside the instrument. «Countermeasure» If the filter is clogged, replace the filter. Show Detail
- No:3675
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» (1) The suction filter on the back of the instrument is clogged. (2) Malfunction inside the instrument. «Countermeasure» (1) If the filter is ... Show Detail
- No:3676
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» (1)D2 lamp malfunction. (2)Malfunction inside the instrument. «Countermeasure» (1) Replace the D2 lamp. Click here for Replace... Show Detail
- No:3677
- Date of Issue:2020/12/02 16:31
- Last Updated:2020/12/07 11:22
- Category: Detector
-
«Main Cause» Lamp compartment fan malfunction «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3678
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Power heat exhaust fan malfunction. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3679
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3680
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3681
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3682
- Date of Issue:2020/12/02 16:31
- Category: Detector
-
«Main Cause» No empty vial in the fraction collector. «Countermeasure» (1) Fraction collector has used up to the final vial. Check the final vial (F.VIAL) s... Show Detail
- No:3683
- Date of Issue:2020/12/02 16:31
- Category: Fraction Collector