• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ HPLC ]

321-330 of Total 943

≪ 33 / 95page ≫
  • Error Code:[0x89D1]

    «Main Cause» Failed to detect light shutter/filter origin. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3674
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D2]

    «Main Cause» PDA Detector: Temperature is increasing inside the instrument. «Countermeasure» If the filter is clogged, replace the filter. Show Detail

    • No:3675
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D3]

    «Main Cause» (1) The suction filter on the back of the instrument is clogged. (2) Malfunction inside the instrument. «Countermeasure» (1) If the filter is ... Show Detail

    • No:3676
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D4]

    «Main Cause» (1)D2 lamp malfunction. (2)Malfunction inside the instrument. «Countermeasure» (1) Replace the D2 lamp. Click here for Replace... Show Detail

    • No:3677
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:22
    • Category: Detector
  • Error Code:[0x89D6]

    «Main Cause» Lamp compartment fan malfunction «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3678
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D7]

    «Main Cause» Power heat exhaust fan malfunction. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3679
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D8]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3680
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89D9]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3681
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89DA]

    «Main Cause» Malfunction inside the instrument. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3682
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8F01]

    «Main Cause» No empty vial in the fraction collector. «Countermeasure» (1) Fraction collector has used up to the final vial. Check the final vial (F.VIAL) s... Show Detail

321-330 of Total 943

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.