  • Error Code:[0x89C3]

    «Main Cause» Probable causes are the D2 lamp connector is not connected, the D2 lamp malfunctioned, or the lighting circuit malfunctioned. «Counter... Show Detail

    • No:3664
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:20
    • Category: Detector
  • Error Code:[0x89C4]

    «Main Cause» Probable causes are the W lamp connector is not connected, or the W lamp malfunctioned. «Countermeasure» Check the W lamp connections. Check t... Show Detail

    • No:3665
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89C6]

    «Main Cause» The lamp cover was not in place at instrument startup. «Countermeasure» Mount the lamp cover, and turn the lamp on again. Show Detail

    • No:3666
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89C7]

    «Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace th... Show Detail

    • No:3667
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89C8]

    «Main Cause» A probable cause is that the cooling fan blast flow is poor, or the cooling fan or heater circuit malfunctioned. «Countermeasure» Ensure at le... Show Detail

    • No:3668
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89C9]

    «Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail

    • No:3669
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89CA]

    «Main Cause» A probable cause is that the D2 lamp heater circuit malfunctioned. «Countermeasure» Click here for Replacement method of Lamp (HPLC... Show Detail

    • No:3670
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:21
    • Category: Detector
  • Error Code:[0x89CB]

    «Main Cause» The D2 lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. «Countermeasure» Connect the D2 lamp conne... Show Detail

    • No:3671
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89CC]

    «Main Cause» The W lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. «Countermeasure» Connect the W lamp connect... Show Detail

    • No:3672
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89CD]

    «Main Cause» A probable cause is that the W lamp malfunctioned. «Countermeasure» Check the total operating time of the lamp, and replace the lamp. If this ... Show Detail

    • No:3673
    • Date of Issue:2020/12/02 16:31
    • Category: Detector