• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ HPLC ]

301-310 of Total 943

≪ 31 / 95page ≫
  • Error Code:[0x8961]

    «Main Cause» Temperature-controlled flow cell malfunctioned, control circuit malfunctioned. «Countermeasure» Turn the power OFF and contact a Shimadzu sale... Show Detail

    • No:3654
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8962]

    «Main Cause» Temperature-controlled flow cell malfunctioned, control circuit malfunctioned. «Countermeasure» Turn the power OFF and contact a Shimadzu sales... Show Detail

    • No:3655
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8964]

    «Main Cause» Cell is not installed or the cell cable is not connected. «Countermeasure» Install the cell and connect the cable correctly. Show Detail

    • No:3656
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8965]

    «Main Cause» (1) An unregistered cell was installed in the instrument. (2) Failed to read cell ID. «Countermeasure» (1) Perform cell registration, exposure... Show Detail

    • No:3657
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8966]

    «Main Cause» The exposure time of the installed cell is not adjusted. «Countermeasure» Set the exposure time. Show Detail

    • No:3658
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8967]

    «Main Cause» The wavelength of the installed cell is not calibrated. «Countermeasure» Perform wavelength calibration. Show Detail

    • No:3659
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8978]

    «Main Cause» Leak sensor is short. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3660
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x8979]

    «Main Cause» Leak sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3661
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x897A]

    «Main Cause» Lamp housing temperature sensor is open. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3662
    • Date of Issue:2020/12/02 16:31
    • Category: Detector
  • Error Code:[0x89C2]

    «Main Cause» D2 lamp malfunctioned. Lighting circuit malfunctioned. «Countermeasure» Replace the D2 lamp. Click here for Replacement method o... Show Detail

    • No:3663
    • Date of Issue:2020/12/02 16:31
    • Last Updated:2020/12/07 11:19
    • Category: Detector

301-310 of Total 943

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.