Error Code:[0x85F1] REMOVE RACK
«Main Cause» The same rack has been set twice. After analysis is performed for this rack, [RC FULL RACK] error is displayed and the scheduled operation is st... Show Detail
- No:3624
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x85F4] ERR DISCONNECT
«Main Cause» The communication between the instrument and the autosampler is cut off due to a communication error. «Countermeasure» Re-insert both connector... Show Detail
- No:3625
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8630] ERR NO RV-* HOME
«Main Cause» Rotation error in an automatic column-switching valve. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3626
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
«Main Cause» Cannot write to oven :Management Device (CMD). «Countermeasure» Once turn off the power, and turn on the power again. If this message still app... Show Detail
- No:3627
- Date of Issue:2020/12/02 16:30
- Category: System Controller , Column Oven
-
Error Code:[0x8636] ERR POST COOLER
«Main Cause» The post-column cooler is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and rem... Show Detail
- No:3628
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x8637] ERR COOLBLK FAN
«Main Cause» The block cooling fan is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and remo... Show Detail
- No:3629
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x8638] ERR DOOR SENS
«Main Cause» The door sensor has failed. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail
- No:3630
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x8650] ERR OVER T.MAX-L
«Main Cause» The column oven temperature of heater L has exceeded the value set for [T.MAX]. «Countermeasure» (1) Lower the column oven temperature set valu... Show Detail
- No:3631
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x81D4] OPEN DRAIN VALVE
«Main Cause» Pump discharge pressure exceeded [P-PMAX] when purge. «Countermeasure» Open drain valve and press [Purge] button. If an error message appears d... Show Detail
- No:3632
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module
-
Error Code:[0x81E2] ERROR OVER HEAT
«Main Cause» Interior temperature has risen to an abnormal level. «Countermeasure» Check to be sure that the internal fan can move, and that the rear exhaus... Show Detail
- No:3633
- Date of Issue:2020/12/02 16:30
- Category: Solvent Delivery Module