FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Error Code:[0x8530] CBM-20A ERROR
«Main Cause» The injection volume exceeds the set value for the maximum volume of injection. «Countermeasure» Reduce the injection volume to a value less th... Show Detail
- No:3584
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8533] ERR COOLER
«Main Cause» There is an error in the sample cooler's cooling unit. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3585
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8534] ERR TEMP SENS DH
«Main Cause» Peltier dehumidifier temperature sensor error. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3586
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8535] ERR TEMP SENS RM
«Main Cause» (1) The room temperature sensor detects an abnormally high or low temperature. (2) Room temperature sensor error. «Countermeasure» (1) If the r... Show Detail
- No:3587
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8537] ERR TEMP SENS L
«Main Cause» Sample cooler L temperature sensor error. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail
- No:3588
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8538] ERR TEMP SENS M
«Main Cause» Sample cooler M temperature sensor error. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail
- No:3589
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8539] ERR TEMP SENS R
«Main Cause» Sample cooler R temperature sensor error. «Countermeasure» Turn off the power and contact a Shimadzu sales or service representative. Show Detail
- No:3590
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8540] NOT ADJUST RACK
«Main Cause» Analysis is started on the sample rack of which rack position has not been adjusted. «Countermeasure» Adjust the rack position. Show Detail
- No:3591
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8541] NOT ADJUSTED-L
«Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate L. «Countermeasure» Adjust the rack position. Show Detail
- No:3592
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8542] NOT ADJUSTED-M
«Main Cause» A sample rack of which rack position has not been adjusted is inserted in rack plate M. «Countermeasure» Adjust the rack position. Show Detail
- No:3593
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
231-240 of Total 943