«Main Cause» attempt was made to use an unregistered rack. «Countermeasure» Set rack number properly. Show Detail
- No:3574
- Date of Issue:2020/12/02 16:30
- Category: Fraction Collector
-
Error Code:[0x0071] ERR SENSOR H
«Main Cause» Thermosensor readings are abnormally high. There may be a short circuit in the sensor. «Countermeasure» Turn power OFF and contact a Shimadzu s... Show Detail
- No:3575
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x0072] ERR SENSOR L
«Main Cause» Thermosensor readings are abnormally low. The sensor may be disconnected. «Countermeasure» Turn power OFF and contact a Shimadzu sales or servi... Show Detail
- No:3576
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x0073] ERR LEAK DETECT
«Main Cause» Density of solvent vapor inside column oven has exceeded leak sensor actuation level. «Countermeasure» (1) Mobile phase has leaked inside the o... Show Detail
- No:3577
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
«Main Cause» Fan speed does not increase to the set value. «Countermeasure» If the cause is obstruction of the fan by tubing or other object, turn the power... Show Detail
- No:3578
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x0076] ERR LEAK SENS
«Main Cause» Leak sensor malfunctions (probably due to broken/disconnected wire). «Countermeasure» Turn power OFF and contact a Shimadzu sales or service re... Show Detail
- No:3579
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x0077] WARNING: LOW SET TEMP
«Main Cause» For CTO-10AC (vp) or CTO-20AC: If the COOLER MODE is "1" or "2", the room temperature is less than -10ºC. When "0" is set, Settings of oven temp... Show Detail
- No:3580
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x0079] COOLER NOT EXIST
«Main Cause» Cooler is not mounted or the cable is snapping. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail
- No:3581
- Date of Issue:2020/12/02 16:30
- Category: Column Oven
-
Error Code:[0x8513] ERR LPV HOME
«Main Cause» Low-pressure valve's rotational position incorrect positions. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service repr... Show Detail
- No:3582
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x8529] SYSTEM ERROR xxx
«Main Cause» There is an error in the Autosampler's internal circuits. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service represen... Show Detail
- No:3583
- Date of Issue:2020/12/02 16:30
- Category: Autosampler