«Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail
- No:3554
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail
- No:3555
- Date of Issue:2020/12/02 16:30
- Category: System Controller
-
«Main Cause» Column oven door was open for too long time while using suppressor. «Countermeasure» Close the door while using suppressor. Show Detail
- No:3556
- Date of Issue:2020/12/02 16:30
- Category: System Controller , Column Oven
-
Error Code:[0x0010] ERR NDLE HOME X
«Main Cause» Move to X (right and left) direction was faulty. «Countermeasure» Check if nothing is stuck in the movable part. If nothing is stuck, contact ... Show Detail
- No:3557
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x0011] ERR NDLE HOME Y
«Main Cause» Move to Y (front and back) direction was faulty. «Countermeasure» Check if nothing is stuck in the movable part. If nothing is stuck, contact ... Show Detail
- No:3558
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x0012] ERR NDLE HOME Z
«Main Cause» Move to Z (up and down) direction was faulty. «Countermeasure» Check if nothing is stuck in the movable part. If nothing is stuck, contact a S... Show Detail
- No:3559
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x0013] ERR HPV HOME
«Main Cause» Rotation of 6-port valve motor was faulty. «Countermeasure» Remedy this problem referring to the following procedure. (1) Check if the rotor ... Show Detail
- No:3560
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x0014] SIL:NO 3-PORT VALVE HOME
«Main Cause» Rotation of 3-way valve motor was faulty. «Countermeasure» Please perform actions using the following procedures. (1) Check that the rotor sea... Show Detail
- No:3561
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x0015] SIL:NO SYRINGE HOME
«Main Cause» Motion of syringe plunger driving motor was faulty. «Countermeasure» The friction between the syringe tip and barrel may have excessively incr... Show Detail
- No:3562
- Date of Issue:2020/12/02 16:30
- Category: Autosampler
-
Error Code:[0x0016] SIL:PFILE PRG ERROR
«Main Cause» There is a line which cannot be executed in the pretreatment program. «Countermeasure» Check set parameter for the line. Show Detail
- No:3563
- Date of Issue:2020/12/02 16:30
- Category: Autosampler