FAQ within [ HPLC ]

191-200 of Total 943

≪ 20 / 95page ≫
  • Error Code:[0x5A70]

    «Main Cause» A broken circuit may have occurred in the cell temperature sensor. «Countermeasure» If the error is not resolved after reconnecting the cell, s... Show Detail

    • No:4106
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A69]

    «Main Cause» Cannot write data to D2 memory. «Countermeasure» Check the D2 lamp connection. If that does not resolved the error, replace the D2 ... Show Detail

    • No:4105
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2020/12/07 11:14
    • Category: Detector
  • Error Code:[0x5A68]

    «Main Cause» Information in the D2 lamp memory could not be read. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail

    • No:4104
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A65]

    «Main Cause» Analysis was started without registering the cell in the main unit. «Countermeasure» Register the cell in the main unit using Utility. Show Detail

    • No:4103
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A64]

    «Main Cause» Analysis was started without connecting the flow cell. «Countermeasure» Connect the cell to the main unit. Show Detail

    • No:4102
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A63]

    «Main Cause» The temperature did not decrease to the setting temperature even after five minutes of temperature control. «Countermeasure» Specify a somewhat... Show Detail

    • No:4101
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A61]

    «Main Cause» The temperature-controlled cell control circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shi... Show Detail

    • No:4100
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A57]

    «Main Cause» The analysis start time received from the system controller is later than the current PDA detector time. «Countermeasure» Restart the PDA detec... Show Detail

    • No:4099
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A56]

    «Main Cause» An instrument IP address conflict occurred with another device. «Countermeasure» Resolve the IP address conflict. If that does not resolve the ... Show Detail

    • No:4098
    • Date of Issue:2020/12/02 16:45
    • Category: Detector
  • Error Code:[0x5A55]

    «Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether PDA detector and system controller... Show Detail

    • No:4097
    • Date of Issue:2020/12/02 16:45
    • Category: Detector