  • Error Code:[0x9100]

    «Main Cause» An sealing ability error occurred in the vacuum flow path of the degassing unit. «Countermeasure» Check the vacuum pressure of the degassing un... Show Detail

  • Error Code:[0x9109]

    «Main Cause» The degassing unit is not connected. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service r... Show Detail

  • Error Code:[0x910C]

    «Main Cause» The backup parameters of the pump are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales o... Show Detail

  • Error Code:[0x910D]

    «Main Cause» The consumable part count data of the pump is corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu... Show Detail

  • Error Code:[0x9413]

    «Main Cause» The set temperature of the column oven is set to 86ºC or higher while the column management device (CMD) is connected. «Countermeasure» Please ... Show Detail

    • No:3548
    • Date of Issue:2020/12/02 16:11
    • Category: Column Oven
  • Error Code:[0x1001]

    «Main Cause» While full-regenerating, activate on or switch cartridge command was operated. «Countermeasure» Do not operate those commands while full regene... Show Detail

  • Error Code:[0x1002]

    «Main Cause» Full-regeneration was operated while HPLC system was activated. «Countermeasure» Do not operate full-regeneration while HPLC system is activat... Show Detail

  • Error Code:[0x1003]

    «Main Cause» The suppressor-related parameters are not correct. «Countermeasure» Check the values and set so that values are correct. Show Detail

  • Error Code:[0x1004]

    «Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail

  • Error Code:[0x1005]

    «Main Cause» Suppressor cartridge current did not become constant. «Countermeasure» The suppressor cartridge may have become degraded. Replace the cartridge. Show Detail